Technical Details:
- Note: All timing measurements made from 10% – 90% amplitude of leading edge under 100k load
- Isolation Method: Digital, using microtransformers
- Isolation Impedance: >1011 Ohms @ 60Hz
- Isolation Barrier Capacitance: <25 pF
- Isolation Voltage: 4000VAC
- Train Time: 1ms – 596.5hrs (24days) at 1ms resolution (continuously adjustable)
- Cycle Time:
- Constant Current Mode: 15μs – 59.6hrs
- Constant Voltage Mode: 4μs – 59.6hrs in 0V-10V range;
- 60μs – 59.6hrs in 10V-125V range (continuously adjustable, on-the-fly adjustable)
- Pulse Delay: 5us – 35min at 1us resolution (on-the-fly adjustable, continuously adjustable)
- Pulse Duration:
- Constant Current Mode: 15μs – 5.9hr
- Constant Voltage Mode: 4μs – 5.9hr in 0V – 10V range;
- 60μs – 5.9hr in 10V – 125V range. (continuously adjustable, on-the-fly adjustable)
- Constant Voltage Ouput: 0 – 125V at 1μA (1Meg load) (continuously adjustable, on-the-fly adjustable)
- Constant Current Ouput: Four ranges, 0 – 10 uA, 0 – 100
- uA, 0 – 1mA, 0 – 10 mA, (continuously adjustable, on-the-fly adjustable)
- Compliance Voltage: +/-125V (10uA and 100uA range) +/-100V (1mA and 10mA ranges)
- Ouput Impedance: <1k
- Leakage Current: <10 nA
- Noise: <1% of range (current or voltage)
- Mounting Options: Tabletop, 4 rubber feet prevent sliding
Ordering Information:
Product number: 55-65-0
Included:
- ESG Unit (55-65-0)
- Stimulus Output Cable
- Rubber Feet
Product Sizes:
- Height: 4.5cm (1.77in)
- Width: 8cm (3.15in)
- Length:16.5cm (6.5in)
- Weight: 2lbs